DISPlay Subsystem R&S FSL
1300.2519.12 6.132 E-11
D
ISPlay[:WINDow<1|2>]:TIME
This command switches on or off the screen display of date and time. The numeric suffix in
WINDow<1| 2> is irrelevant.
This command is available from firmware version 1.10.
Parameter
ON | OFF
Example
DISP:TIME ON
Characteristics
*RST value: OFF
SCPI: device-specific
Mode
all
DISPlay[:WINDow<1|2>]:TRACe<1...6>[:STATe]
This command switches on or off the display of the corresponding trace.
The numeric suffixes <1|2> are irrelevant.
Parameter
ON | OFF
Example
DISP:TRAC3 ON
Characteristics
*RST value: ON for TRACe1, OFF for TRACe2 to 6
SCPI: conform
Mode
all
DISPlay[:WINDow<1|2>]:TRACe<1...6>:MODE
This command defines the type of display and the evaluation of the traces. WRITE corresponds
to the Clr/Write mode of manual operation. The trace is switched off (= BLANK in manual
operation) with DISPlay[:WINDow<1|2>]:TRACe<1...6>[:STATe].
The number of measurements for AVERage, MAXHold and MINHold is defined with the
[SENSe<1|2>:]AVERage:COUNt or [SENSe<1|2>:]SWEep:COUNt commands. It should be
noted that synchronization to the end of the indicated number of measurements is only possible
in single sweep mode.
If calculation of average values is active, selection between logarithmic and linear averaging is
possible. For more detail see [SENSe<1|2>:]AVERage:TYPE command.
The numeric suffixes <1|2> are irrelevant.
Parameter
WRITe | VIEW | AVERage | MAXHold | MINHold |RMS
For details on trace modes refer to chapter "Instrument Functions", section "Trace mode
overview".