DISPlay Subsystem R&S FSL
1300.2519.12 6.134 E-11
Characteristics
*RST value: LOG
SCPI: conform
Mode
A-F
DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y:SPACing
This command selects the scaling for the level display range.
The numeric suffixes <1|2> and <1...6> are irrelevant.
Parameter
LOGarithmic Selects logarithmic scaling.
LINear Selects linear scaling in %.
LDB Selects linear scaling in dB.
Example
DISP:TRAC:Y:SPAC LIN
Characteristics
*RST value: LOGarithmic
SCPI: conform
Mode
A, ADEMOD, CATV
DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]
This command defines the display range of the y–axis (level axis) with logarithmic scaling
(DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y:SPACing).
For linear scaling, the display range is fixed and cannot be modified.
The numeric suffixes <1|2> and <1...6> are irrelevant.
Parameter
10 dB to 200 dB or value in Hz
Example
DISP:TRAC:Y 110dB
Characteristics
*RST value: 100dB
SCPI: device–specific
Mode
all