MMEMory Subsystem R&S FSL
1300.2519.12 6.164 E-11
MMEMory:SELect[:ITEM]:FINal
This command includes the final measurement data in the list of partial datasets of a device
setting to be stored/loaded.
Parameter
ON | OFF
Example
"MMEM:SEL:FIN ON"
inserts the final measurement data in the list of partial data sets
Characteristics:
*RST value: ON
SCPI: device-specific
Mode
R, A
MMEMory:SELect[:ITEM]:DEFault
This command includes the default list in the settings file to be stored/loaded. For details on
hardware settings refer to the MMEMory:SELect[:ITEM]:HWSettings command.
Example
MMEM:SEL:DEFault
Characteristics
*RST value: -
SCPI: device-specific
Mode
all
MMEMory:SELect[:ITEM]:HWSettings
This command includes the hardware settings in the settings file to be stored/loaded. The
hardware settings include:
• current configuration of general device parameters (general setup)
• current setting of the measurement hardware including markers
• activated limit lines:
A settings file may include 8 limit lines at maximum. This number includes the activated limit
lines and, if available, the de-activated limit lines last used.
Therefore the combination of the non-activated restored limit lines depends on the sequence of
use with the MMEMory:LOAD:STATe command.
• user-defined color setting
• configuration for the hardcopy output
• tracking generator settings (models 13, 16 and 28)
• correction data for source calibration (models 13, 16 and 28)
Parameter
ON | OFF