SENSe Subsystem R&S FSL
1300.2519.12 6.180 E-11
[
SENSe<1|2>:]CORRection:COLLect[:ACQuire]
When the tracking generator is active, this command determines the type of result acquisition
for the normalization reference measurement and starts the measurement selected:
To obtain a correct reference measurement, a complete sweep with synchronization to the end
of the sweep must have been carried out. This is only possible in the single sweep mode.
The numeric suffixes <1|2> are not relevant.
This command is an "event" and therefore has no *RST value and no query.
Parameter
THRough "TRANsmission" mode calibration with direct connection
between tracking generator and
device input
"REFLection" mode calibration with short circuit at the
input
OPEN only allowed in "REFLection" mode calibration with open input
Example
INIT:CONT OFF
Selects single sweep operation
CORR:COLL THR;*WAI
Starts the measurement of reference data using direct connection between generator and
device input and waits for the sweep end.
Characteristics
*RST value: –
SCPI: conform
Mode
A
[SENSe<1|2>:]CORRection:EGAin:INPut[:MAGNitude]
This command makes an external gain known to the analyzer, which will take it into account
during the display of measurement results. With this function the gain of an antenna or of an
external preamplifier can be taken into account for the measurement values.
The numeric suffixes <1|2> are not relevant.
Parameter
–200...200dB
Example
CORR:EGA:INP 10DB
Takes 10 dB external gain into account.
Characteristics
*RST value: 0 dB
SCPI: device–specific
Mode
A