R&S FSL SENSe Subsystem
1300.2519.12 6.181 E-11
[SENSe<1|2>:]CORRection:METHod
This command selects the type of measurement with active tracking generator
(transmission/reflection).
The numeric suffixes <1|2> are not relevant.
Parameter
TRANsmission | REFLection
Example
CORR:METH TRAN
Sets the type of measurement to "transmission".
Characteristics
*RST value: TRANsmission
SCPI: device–specific
Mode
A
[SENSe<1|2>:]CORRection:RECall
This command restores the instrument setting that was applied to the measurement of the
reference data, provided that the tracking generator is active.
The numeric suffixes <1|2> are not relevant.
This command is an event and therefore has no *RST value and no query.
Example
CORR:REC
Characteristics
*RST value: –
SCPI: conform
Mode
A
[SENSe<1|2>:]CORRection:TRANsducer[:STATe]
This command switches the selected transducer factor on or off.
Note: Prior to this command, the [SENSe<1|2>:]CORRection:TRANsducer:SELect
command must be sent.
This command is available from firmware version 1.10.
Parameter
ON | OFF
Example
CORR:TRAN ON
Characteristics
*RST value: OFF
SCPI: device-specific
Mode A