EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 941

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSe Subsystem R&S FSL
1300.2519.12 6.182 E-11
[SENSe<1|2>:]CORRection:TRANsducer:ADJust:RLEVel[:STATe]
This command switches the automatic adjustment of the reference level to the selected
transducer factor on or off. For details refer to chapter 4, section "Instrument Setup and
Interface Configuration - SETUP Key", Ref Level Adjust Man/Auto softkey.
Note: Prior to this command, the [SENSe<1|2>:]CORRection:TRANsducer:SELect
command must be sent.
This command is available from firmware version 1.10.
Parameter
ON | OFF
Example
CORR:TRAN:ADJ:RLEV ON
Characteristics
*RST value: OFF
SCPI: device-specific
Mode
A
[SENSe<1|2>:]CORRection:TRANsducer:COMMent
This command defines the comment for the selected transducer factor.
Note: Prior to this command, the [SENSe<1|2>:]CORRection:TRANsducer:SELect
command must be sent.
This command is available from firmware version 1.10.
Parameter
<string>
Example
CORR:TRAN:COMM 'FACTOR FOR ANTENNA'
Characteristics
*RST value: (empty comment)
SCPI: device-specific
Mode A
[SENSe<1|2>:]CORRection:TRANsducer:DATA
This command defines the reference values of the transducer factor selected. These values are
entered as a sequence of frequency/level pairs. The frequencies must be sent in ascending
order.
Note: Prior to this command, the [SENSe<1|2>:]CORRection:TRANsducer:SELect
command must be sent. The level values are sent as dimensionless numbers; the unit
is specified by means of the [SENSe<1|2>:]CORRection:TRANsducer:UNIT
command.
This command is available from firmware version 1.10.
Parameter
<freq>,<level>