R&S FSL SENSe Subsystem
1300.2519.12 6.183 E-11
Example
SENS:CORR:TRAN:SEL 'TRAN_1'
Selects the transducer factor
CORR:TRAN:UNIT 'DB'
Selects the unit for the transducer.
CORR:TRAN:DATA 1MHZ,-30,2MHZ,-40
Defines the frequency/level pairs 1 MHz, -30 dB and 2 MHz, -40 dB.
Characteristics
*RST value: -
SCPI: device-specific
Mode
A
[SENSe<1|2>:]CORRection:TRANsducer:DELete
This command deletes the selected transducer factor.
This command is an event and therefore has no *RST value.
Note: Prior to this command, the [SENSe<1|2>:]CORRection:TRANsducer:SELect
command must be sent.
This command is available from firmware version 1.10.
Example
CORR:TRAN:DEL
Characteristics
*RST value: -
SCPI: device-specific
Mode
A
[SENSe<1|2>:]CORRection:TRANsducer:SCALing
This command defines whether the frequency scaling of the transducer factor is linear or
logarithmic.
Note: Prior to this command, the [SENSe<1|2>:]CORRection:TRANsducer:SELect
command must be sent.
This command is available from firmware version 1.10.
Parameter
LINear | LOGarithmic
Example
CORR:TRAN:SCAL LOG
Characteristics
*RST value: LINear
SCPI: device-specific
Mode
A