R&S FSL SENSe Subsystem
1300.2519.12 6.179 E-11
SENSe:CORRection Subsystem (Models with tracking generator)
This subsystem controls calibration and normalization during operation with the tracking generator.
Commands of the SENSe:CORRection Subsystem
[SENSe<1|2>:]CORRection[:STATe]
[SENSe<1|2>:]CORRection:COLLect[:ACQuire]
[SENSe<1|2>:]CORRection:EGAin:INPut[:MAGNitude]
[SENSe<1|2>:]CORRection:METHod
[SENSe<1|2>:]CORRection:RECall
[SENSe<1|2>:]CORRection:TRANsducer[:STATe]
[SENSe<1|2>:]CORRection:TRANsducer:ADJust:RLEVel[:STATe]
[SENSe<1|2>:]CORRection:TRANsducer:COMMent
[SENSe<1|2>:]CORRection:TRANsducer:DATA
[SENSe<1|2>:]CORRection:TRANsducer:DELete
[SENSe<1|2>:]CORRection:TRANsducer:SCALing
[SENSe<1|2>:]CORRection:TRANsducer:SELect
[SENSe<1|2>:]CORRection:TRANsducer:UNIT
[SENSe<1|2>:]CORRection:TRANsducer:VIEW
[SENSe<1|2>:]CORRection[:STATe]
This command activates/deactivates the normalization of the measurement results provided that
the tracking generator is active. The command is available only after acquisition of a reference
trace for the selected type of measurement (transmission/reflection, see
[SENSe<1|2>:]CORRection:COLLect[:ACQuire] command).
The numeric suffixes <1|2> are not relevant.
Parameter
ON | OFF
Example
CORR ON
Activates normalization.
Characteristics
*RST value: OFF
SCPI: conform
Mode
A