SENSe Subsystem R&S FSL
1300.2519.12 6.186 E-11
SENSe:DETector Subsystem
The SENSe:DETector subsystem controls the acquisition of measurement data via the selection of the
detector for the corresponding trace.
Commands of the SENSe:DETector Subsystem
[SENSe<1|2>:]DETector<1...6>[:FUNCtion]
[SENSe<1|2>:]DETector<1...6>[:FUNCtion]:AUTO
[SENSe<1|2>:]DETector<1...6>[:FUNCtion]
This command switches on the detector for the data acquisition in the selected trace.
The trace is indicated as numeric suffix in DETector.
The numeric suffixes <1|2> are not relevant.
Parameter
APEak | NEGative | POSitive | SAMPle | RMS | AVERage | QPEak
For details on detectors refer to chapter "Instrument Functions", section "Detector overview".
Example
DET POS
Sets the detector to "positive peak".
Characteristics
*RST value: APEak
SCPI: conform
Mode
A, CATV
[SENSe<1|2>:]DETector<1...6>[:FUNCtion]:AUTO
This command either couples the detector to the current trace setting or turns coupling off. The
trace is selected by the numeric suffix at DETector.
The numeric suffixes <1|2> are not relevant.
Parameter
ON | OFF
Example
DET:AUTO OFF
Characteristics
*RST value: ON
SCPI: conform
Mode
A, CATV