SENSe Subsystem R&S FSL
1300.2519.12 6.204 E-11
[SENSe<1|2>:]FREQuency:STARt
This command defines the start frequency of the analyzer. This command is only available with
span > 0.
The numeric suffixes <1|2> are not relevant.
Parameter
0 to f
m
ax
f
m
ax
is specified in the data sheet. To help analyze signals located at the end of the frequency
range, for R&S FSL models with an upper frequency limit of 6 GHz or less, the f
max
value is
extended by 0.05 GHz for direct entry. The preset and maximum values remain unchanged.
Example
FREQ:STAR 20MHz
Characteristics
*RST value: 0
SCPI: conform
Mode
A–F, CATV, CDMA, EVDO
[SENSe<1|2>:]FREQuency:STOP
This command defines the stop frequency of the analyzer. This command is only available with
span > 0.
The numeric suffixes <1|2> are not relevant.
Parameter
0 to f
max
f
max
is specified in the data sheet. To help analyze signals located at the end of the frequency
range, for R&S FSL models with an upper frequency limit of 6 GHz or less, the f
max
value is
extended by 0.05 GHz for direct entry. For the R&S FSL18 model, the f
max
value is extended to
20 GHz. The preset and maximum values remain unchanged.
For the R&S FSL18, the query returns a maximum frequency of 20 GHz.
Example
FREQ:STOP 2000MHz
Characteristics
*RST value: f
max
SCPI: conform
Mode
A–F, CATV, CDMA, EVDO

Table of Contents

Related product manuals