SENSe Subsystem R&S FSL
1300.2519.12 6.208 E-11
[SENSe<1|2>:]LIST:POWer:RESult?
This command queries the result of a previous list measurement as configured and initiated with
[SENSe<1|2>:]LIST:POWer[:SEQuence]. The measured results are output in a list of
floating point values separated by commas. The unit of the results depends on the setting made
with the CALCulate<1|2>:UNIT:POWer command.
This command may be used to obtain measurement results in an asynchronous way, using the
service request mechanism for synchronization with the end of the measurement.
If no measurement results are available, the command will return a query error.
The numeric suffixes <1|2> are not relevant.
Example
*ESE 1
*SRE 32
Configuration of the status reporting system for the generation of an SRQ on operation
complete
SENSe:LIST:POWer
935.2MHz,–20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,–20dBm,10dB,10dB,NORM,30kHz,100kHz,434us,0,
935.6MHz,–20dBm,10dB,20dB,NORM,30kHz,100kHz,434us,0;
*OPC
Configuring and starting the measurement
...
Further actions of the control computer during measurement
On SRQ:
SENSe:LIST:POWer:RESult?
Response to service request
Characteristics
*RST value: –
SCPI: device–specific
Mode
A–F, A–T
[SENSe<1|2>:]LIST:POWer:SET
This command defines the constant settings for the list during multiple power measurement.
Parameters <PEAK meas>, <RMS meas> and <AVG meas> define, which measurements are
to be performed at the same time at the frequency point. Correspondingly, one, two or three
results per frequency point are returned for the [SENSe<1|2>:]LIST:POWer[:SEQuence]
command. If all three parameters are set to OFF, the command generates an execution error.
The numeric suffixes <1|2> are not relevant.
Parameter
<PEAK meas> ON: activates the measurement of the peak power (peak detector)
OFF: deactivates the measurement of the peak power
<RMS meas> ON: activates the measurement of the RMS power (RMS detector)
OFF: deactivates the measurement of the RMS power
<AVG meas> ON: activates the measurement of the average power (average
detector)
OFF: deactivates the measurement of the average power
<trigger mode> Selection of the trigger source used for the list measurement