R&S FSL SENSe Subsystem
1300.2519.12 6.209 E-11
Possible values: IMMediate | EXTernal | VIDeo | IFPower
<trigger slope> Used trigger slope
Possible values: POSitive | NEGative
<trigger offset> Offset between the detection of the trigger signal and the start of the
measurement at the next frequency point.
R
ange of values: 0 s, 125 ns to 100s
<gate length> Gate length with gated sweep
Range of values: 0 s, 125 ns to 100s
The value 0 s deactivates the use of gated trigger; other values
activate the gated trigger function.
Values <> 0 s are only possible if <trigger mode> is different from
IMMediate. Otherwise, an execution error is triggered.
Return values
The query command returns a list of comma–separated values (CSV) of the settings, i.e.
ON,ON,ON,IMM,POS,0,0
if the configuration has been set with the command
SENSe:LIST:POWer:SET ON,ON,ON,IMM,POS,0,0
Example
SENSe:LIST:POWer:SET ON,OFF,OFF,EXT,POS,10US,434US
Characteristics
*RST value: ON,OFF,OFF,IMM,POS,0S,0S
SCPI: device–specific
Mode
A–F, A–T
[SENSe<1|2>:]LIST:POWer:STATe
This command deactivates the list measurement.
The numeric suffixes <1|2> are not relevant.
Parameter
OFF
Example
SENSe:LIST:POWer:STATe OFF
Characteristics
*RST value: –
SCPI: device–specific
Mode
A–F, A–T