R&S FSL SENSe Subsystem
1300.2519.12 6.221 E-11
[SENSe<1|2>:]POWer:ACHannel:BANDwidth|BWIDth[:CHANnel]
This command sets the channel bandwidth of the radio communication system. The bandwidths
of adjacent channels are not influenced by this modification.
With [SENSe<1|2>:]POWer:HSPeed set to ON, steep–edged channel filters are available.
For further information on filters refer to chapter "Instrument Functions", section "List of
available RRC and channel filters".
The numeric suffixes <1|2> are not relevant.
Parameter
100 Hz to 1000 MHz
Example
POW:ACH:BWID 30kHz
Sets the bandwidth of the TX channel to 30 kHz.
Characteristics
*RST value: 14 kHz
SCPI: device–specific
Mode
A–F
[SENSe<1|2>:]POWer:ACHannel:BANDwidth|BWIDth:ACHannel
This command defines the channel bandwidth of the adjacent channel of the radio transmission
system. If the bandwidth of the adjacent channel is changed, the bandwidths of all alternate
adjacent channels are automatically set to the same value.
With [SENSe<1|2>:]POWer:HSPeed set to ON, steep–edged channel filters are available.
For further information on filters refer to chapter "Instrument Functions", section "List of
available RRC and channel filters".
The numeric suffixes <1|2> are not relevant.
Parameter
100 Hz to 1000 MHz
Example
POW:ACH:BWID:ACH 30kHz
Sets the bandwidth of all adjacent channels to 30 kHz.