4. TSM Service Program 1/2010 4 - 162
BA-TE-DE08W M.KAY Dialog+ SW9xx_SM_Chapter 4-3-1_1_2010.doc/pdf <110301> yymmdd B. Braun Avitum AG
Dialog
+
SW 9.xx
4.8.3.21 Calibration Dialysate Nonreturn
Valve RVDA
You can calibrate the dialysate nonreturn valve RVDA.
DDE
FBK2
FBK1
RVDA
VABK2
VDABK2
VABK1
VDABK1
BK2
BK1
VEBK2
VDEBK2
VEBK1
VDEBK1
MSBK2
MSBK1
VZ
RVFPE
FPE
8
Loop-In Point
for Calibration
Fig.: Calibration RVDA/Loop-In Point
Fig.: UF Sub-Rack with FBK1
Fig.: Water Sub-Rack with RVDA
FBK1
EP
FEP
RVDA
Loop-In Point
for Calibration
WAB