M Mode
6.2.3 Sweep Speed
On the status area you will find the [Speed] key. review: M Main Menu (chapter 6.1)
By touching up or down four different sweep speeds can be selected.
1 = 3,5 cm/s
2 = 5,0 cm/s
3 = 7,5 cm/s
4 = 10,0 cm/s (in relation to the system’s monitor)
Remark:
The Speed is only adjustable in write mode.
6.2.4 Invert
This function inverts the M mode trace from Up to Down in the M mode display area.
key unlit : M mode display normal
key lit: M mode display inverted
Remark:
The Invert function is only available with endo-vaginal probes.
6.2.5 Frequency
It is the same as the 2D mode frequency.
review: Receiver Frequency Range
(chapter 5.8)
6.2.6 TGC Slide Controls
The [TGC]- and the [Transmit Power] settings are the same for the M trace and the 2D image.
TGC setting review: TGC Slider Controls
(chapter 5.5)
Transmit Power review: Transmit Power
(chapter 5.6)
6.2.7 M Mode Depth
The function is the same as 2D mode depth
review: 2D Mode Depth
(chapter 5.2)
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0 6-5