Calculations and Patient Reports
9. Move the second cursor to the anterior IVSs signal and press the right or left trackball key.
The mark is fixed and the next cursor will appear.
10. Move the cursor to LVDs and press the right or left trackball key.
The mark is fixed and the next cursor will appear.
11. Move the cursor to LVPWs and press the right or left trackball key.
The mark will be fixed and the measurement of the Left Ventricle is completed.
Note:
Additionally also [RVDd] and [Heart Rate] can be measured.
If the Heart Rate is measured, also the Cardiac Output is calculated.
review: Cardiac Report
(chapter 14.8)
IVSd: Interventricular Septum - diastolic
LVDd: Left Ventricle Diameter - diastolic
LVPWd: Left Ventricular Posterior Wall - diastolic
IVSs: Interventricular Septum - systolic
LVDs: Left Ventricle Diameter -systolic
LVPWs: Left Ventricular Posterior Wall - systolic
14.7.4.6.2 To measure the Items One by One
To measure the items such as IVSd, LVDd, LVPWd, IVSs, LVDs, LVPWs and RVDd.
1. Press the [Calc] key.
The Cardiac calculations menu appears in the menu area.
2. Select [LV M-Mode] and select the appropriate item.
3. Perform the measurement using trackball and the right or left
trackball key [Set].
Voluson
®
730Pro - Basic User Manual
14-32 105831 Rev. 0