MAN ECG preamplifier
22. MAN ECG preamplifier
22.1 Description
The ECG module consists of a ECG preamplifier type MAN 6 (hardware part) and a patient
connection cable (KENDALL medizinische Erzeugnisse, Code 8/P93/07-01, Ref 3227.0701.00).
The connector for the patient connection cable is on the front cover of the hardware section, which
is placed in a drive-slot in the front of the ultrasound machine.
The ECG preamplifier type MAN6 is used for acquiring an ECG signal to be displayed with the
ultrasound image. The ECG preamplifier must not be used for ECG diagnostics. It is not intended
for use as a cardiac monitor.
The signal input of the ECG preamplifier is equipped with a protection against high voltages used
for defibrillation (Type BF).
The ECG preamplifier is connected to a connector on the rear panel of Voluson
®
730Pro.
review also: Technical Data/Information: MAN ECG preamplifier