Calculations and Patient Reports
14.9.3.4 Heart Rate
The Heart Rate is to be calculated in the M or Spectral-Doppler mode.
1. Press the [Calc] key on the control panel.
The Vascular calculations menu appears in the menu area.
2. Select the item Heart Rate [HR] in the menu.
3. When the vertical line appears on the screen, move it to the start
point of the period using the trackball and the right or left
trackball [Set] key. The second line appears.
4. Select the number of heart rate cycles necessary for
measurement with the digipot.
5. If necessary, adjust the [Angle] and the [Baseline].
6. Move the second line to the selected number of heart rate cycles to be measured.
Then press the right or the left trackball [Set] key. The Heart Rate is displayed.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0 14-47