Utilities
12.5 Display of Biopsy Guideline
WARNING!
• The biopsy lines must be programmed once by the service personnel or by the user.
The procedure must be repeated if probes and/or biopsy guides are exchanged!
• Before performing a biopsy make sure that the displayed biopsy line coincides with the
needle track (check in a bowl filled with approx. 47°C warm water).
• Please read the "Instructions for safe Use" in Probes and Biopsy
(chapter 20).
Operation:
Main Menu has to be active :
Press the [Utilities] key.
Select the [Biopsy] item.
The Biopsy guideline is switched ON if the Biopsy key is highlighted.
To switch OFF the Biopsy guideline select the [Biopsy] item again.
It toggles between ON and OFF.
To program a Biopsy guideline review: To program a Biopsy Line
(chapter 19.1)
Remarks
:
* For handling (sterilization etc. ) and operation of the biopsy guides
review: Probes and Biopsy
(chapter 20).
* For each probe one biopsy guideline is programmable.
Voluson
®
730Pro - Basic User Manual
105831 Rev. 0 12-11