MAN ECG preamplifier
22.2 Handling
The ECG function is switched on and off by pressing the [ECG] key on the control panel
of the ultrasound machine.
• Position, speed and amplitude of the displayed ECG strip can be altered in the “ECG menu” shown
in the menu or status area of the ultrasound machine.
• The patient cable shall always be connected to the ECG preamplifier.
• With the patient cable belonging to the ECG-preamplifier, only electrodes for push-button
connection can be used. Depending on requirements, commercially available extremity clamp
electrodes together with conductive gel or commercially available pre-jelled adhesive electrodes
can be used, preferably the latter should be used.
• With standard setting of the electrodes (red = right arm, yellow = left arm, black = left leg) lead I is
displayed. Other electrode arrangements may be necessary (lead II, III), if amplitude supplied by
lead I is too small.
Voluson
®
730Pro - Operation Manual
105831 Rev. 0 22-3