MAN ECG preamplifier
22.2 Handling
The ECG function is switched on and off by pressing the [ECG] key on the control panel
of the ultrasound machine.
Position, speed and amplitude of the displayed ECG strip can be altered in the “ECG menu” shown
in the menu or status area of the ultrasound machine.
The patient cable shall always be connected to the ECG preamplifier.
With the patient cable belonging to the ECG-preamplifier, only electrodes for push-button
connection can be used. Depending on requirements, commercially available extremity clamp
electrodes together with conductive gel or commercially available pre-jelled adhesive electrodes
can be used, preferably the latter should be used.
With standard setting of the electrodes (red = right arm, yellow = left arm, black = left leg) lead I is
displayed. Other electrode arrangements may be necessary (lead II, III), if amplitude supplied by
lead I is too small.
Voluson
®
730Pro - Operation Manual
105831 Rev. 0 22-3