Probes and Biopsy
20.14.3 Programming and Display
Immerse the transducer with the mounted biopsy needle guide in a vessel filled with warm water
(approx. 47°C or 117°F).
Insert the needle into the needle guide, until a needle echo can be seen in the ultrasound image.
Set the power setting and the gain setting to the minimum necessary for a good result.
For further instructions review: To program a Biopsy Line
(chapter 19.1).
Biopsy Probe Capabilities and Features Illustration
PEC42
IC5-9
Needle diameters:
< 1.8 mm
Material:
Stainless Steel
Sterilization with autoclave
possible!
PEC63
RIC5-9
Needle diameters:
< 1 8 mm
Material:
Stainless Steel
Sterilization with autoclave
possible!
PEC64
SP6-12
Needle diameters:
< 1 mm
< 1.4 mm
< 2.2 mm
Material:
Stainless Steel
Sterilization with autoclave
possible!
PEC65
RAB2-5
RAB4-8P
Needle diameters:
< 1 mm
< 1.4 mm
< 2.2 mm
Material:
Stainless Steel
Sterilization with autoclave
possible!
Voluson
®
730Pro - Basic User Manual
20-26 105831 Rev. 0