MAN ECG preamplifier
22.5 ECG Display
This function inserts an ECG line in the display of the image.
Condition:
ECG module (ECG preamplifier) is connected to the system.
Operation:
1. Press the [ECG] key on the control panel to switch on/off the ECG line.
The status area shows additional functions for the ECG Display.
2. Adjust the transmission gain of the ECG preamplifier signal (0, 1, 2, 3).
3. Select ECG velocity (0, 1, 2, 3).
4. Set the vertical position on the monitor.
5. Adjust ECG amplitude ( 0 to 100 in 10 steps).
6. Return to the main menu.
The ECG function remains active.
7. Freeze the image.
The most recent information is always on the right edge of the image.
When moving the trackball a indicator (small vertical line) is inserted on the
ECG curve and indicates the temporal position of the 2D image in relation to the
recorded ECG-line. In this manner, e.g., diastolic or systolic phase of the 2D mode
image can be set (without ECG trigger).
Voluson
®
730Pro - Operation Manual
105831 Rev. 0 22-5