System Setup
17.3.8.1 To Specify a DICOM Address
Add: To add a new DICOM node click on the [Add] button.
Edit: To edit or view data of a DICOM node, select one by its alias name, then click the
[Edit] button.
Delete: To delete a DICOM node click the [Delete] button.
After clicking the [Add] or [Edit] button the “DICOM Device Setup” window appears. (e.g., PRINT)
To specify a DICOM address to define the following fields:
Services: 1. Select [STORE] to send images to a DICOM server ( e.g., PC.....).
2. Select [PRINT] to send images to a DICOM printer.
3. Select [WORKLIST] to retrieve Patient Information (Name, ID, Birth,...)
from external Worklist server (e.g., HIS -
Hospital Information System / RIS).
4. Select [REPORT] to send the Patient report to a PC.
By selecting [PRINT] the Printer Setup fields are available to adjust the printer configuration.
AE-Title: Application Entity Title of the remote DICOM application.
Alias: Enter a nickname for each DICOM node to make it easier to handle various
nodes. Use any name, but do not insert space characters.
IP-Address: Enter the hostname or IP-address of the DICOM node.
e.g., any.dicom.server.net
Port Number: Enter the port number of the DICOM node.
e.g., 104