Volume Mode
8. Select the Quality (low, mid1, mid2, high1, high2).
This function changes the line density against acquisition speed.
Low: Fast speed / low scan density
This mode is selected only in case of expected movement artifacts.
A loss of volume resolution will result
Mid2: Standard VOL scan / medium scan density
High1: Slow speed/ high scan density
9. To start 3D acquisition press the [Freeze] key, respectively the right
trackball key (Start -> displayed in Status area of the trackball).
The volume acquisition starts.
During the acquisition of a 3D volume, the unit displays only the volume box area.
After acquisition the unit changes to read mode. review: After the 3D Acquisition
(chapter 11.2.2)
To stop the acquisition, press the [Exit] key on the control panel.
The acquisition stops and the “3D Pre” menu appears again.
Note:
The recorded information will be cleared!