Volume Mode
8. Select the Quality (low, mid1, mid2, high1, high2).
This function changes the line density against acquisition speed.
Low: Fast speed / low scan density
This mode is selected only in case of expected movement artifacts.
A loss of volume resolution will result
Mid2: Standard VOL scan / medium scan density
High1: Slow speed/ high scan density
9. To start 3D acquisition press the [Freeze] key, or the right trackball key
(Start displayed in Status area of the trackball).
The volume acquisition begins.
During the Acquisition of a 3D volume the unit displays only the volume box area.
To stop the Acquisition, press the [Exit] key on the control panel.
The Acquisition stops and the “3D Pre” menu appears again.
Note:
The recorded information will be cleared!
To perform a 3D Acquisition During Active High Resolution Zoom
review chapter 11.2.1
Condition to activate 3D Image Rendering:
1. A “Static 3D Sectional Planes” volume scan is performed and in read mode.
Therefore select the [3D Rendering] item after 3D Sectional Planes
acquisition.
2. A volume scan is performed with “Static 3D Render”.
Voluson
®
730Pro - Basic User Manual
11-32 105831 Rev. 0