EasyManua.ls Logo

MBRAUN LABmaster pro sp - Page 101

MBRAUN LABmaster pro sp
302 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 Transport, installation and commissioning
Commissioning
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
4 - 33
Performing a manual purge
1 Purge gas source 2 Pressure reducing valve, 3.0 bar
(0.3 MPa)
3 Reinforced hose 4 Open flange
5 Inert gas exhaust into the glovebox 6 Mixed ambient air and inert gas es-
cape
In a two-glovebox system, each glovebox must be purged separately. Both boxes
must have an open connection to the circulation pipework.
Open freezers or equipment with areas that may be protected by covers.
Set up purge gas source (working gas) with pressure reducing valve.
Connect reinforced hose to purge gas source.
Open one KF40 “blind flange” on glovebox.
Feed one end of the reinforced hose through the open flange into the gloves.
Set the pressure-reducing valve on the purge gas source to 3.0 bar (0.3 MPa) and
open the valve.
ð Air and purge gas escapes through the flange opening.
Pick up the reinforced hose with a glove, and purge the glovebox interior from top to
bottom using a circular motion.
Carefully purge corners, edges and box fittings.
The purge time should be about 45 min per 1m
3
of glovebox volume to achieve an appro-
priate gas displacement with an oxygen concentration of less than 100 ppm (V).

Table of Contents

Related product manuals