CALCulate Subsystem (SPECM, K14) R&S FSL
1300.2519.12 6.394 E-11
CALC:DELT4:SPEC:Y:MAX:NEXT
Positions delta marker 4 on the next maximum below or above its current position.
Characteristics
RST value: –
SCPI: device–specific
Mode
SPECM
CALCulate<1|2>:DELTamarker<1...4>:SPECtrogram:Y:MINimum[:PEAK]
This command positions the specified delta marker on the current minimum in y–axis direction
(constant x–axis value). The corresponding search in x direction is performed via the
CALCulate<1|2>:DELTamarker<1...4>:MINimum[:PEAK] command.
This command is an event and therefore has no *RST value and no query.
The numeric suffixes <1|2> are irrelevant for this command.
Example
CALC:DELT4 ON
Switches on delta marker 4 or switches to delta marker mode.
CALC:DELT4:SPEC:Y:MIN
Positions delta marker 4 on the current minimum in y–axis direction.
Characteristics
RST value: –
SCPI: device–specific
Mode
SPECM
CALCulate<1|2>:DELTamarker<1...4>:SPECtrogram:Y:MINimum:ABOVe
This command positions the specified delta marker on the next minimum above its current
position (constant x–axis value). The corresponding search in x direction is performed via the
CALCulate<1|2>:DELTamarker<1...4>:MINimum:LEFT and
CALCulate<1|2>:DELTamarker<1...4>:MINimum:RIGHt commands.
This command is an event and therefore has no *RST value and no query.
The numeric suffixes <1|2> are irrelevant for this command.
Example
CALC:DELT2 ON
Switches on delta marker 2 or switches to delta marker mode.
CALC:DELT2:SPEC:Y:MIN:ABOV
Positions delta marker 2 on the next minimum above its current position.
Characteristics
RST value: –
SCPI: device–specific
Mode