R&S FSL CALCulate Subsystem (CATV, K20)
1300.2519.12 6.419 E-11
CALCulate<1|2>:ATV:LIMit:HUM:HUM:LOWer
This command defines the lower limit for the hum values in dB. If the wrong unit is set, the
command produces an execution error. The unit is set via the
CALCulate<1|2>:ATV:UNIT:POWer:HUM command.
The numeric suffixes <1|2> are irrelevant.
Parameter
–60 to 60 dB
Example
CALC:ATV:UNIT:POW:HUM DB
Sets the unit for the Hum measurement values to dB.
CALC:ATV:LIM:HUM:HUM:LOW 10
Sets the lower limit for the hum values to 10 dB.
Characteristics
RST value: 40 dB
SCPI: device–specific
Mode
CATV
CALCulate<1|2>:ATV:LIMit:HUM:HUM:UPPer
This command defines the upper limit for the hum values in %. If the wrong unit is set, the
command produces an execution error. The unit is set via the
CALCulate<1|2>:ATV:UNIT:POWer:HUM command.
The numeric suffixes <1|2> are irrelevant.
Parameter
0.1 to 100.000%
Example
CALC:ATV:UNIT:POW:HUM PCT
Sets the unit for the Hum measurement values to percent.
CALC:ATV:LIM:HUM:HUM:UPP 6
Sets the upper limit of the hum values to 6%.
Characteristics
RST value: 1%
SCPI: device–specific
Mode
CATV

Table of Contents