CALCulate Subsystem (Noise Figure, K30) R&S FSL
1300.2519.12 6.530 E-11
CALCulate:MARKer:TRACe
This command assigns the marker1 to the indicated trace.
This command is available from firmware version 1.60.
Parameter
NOISe | GAIN
E
xample
CALC:MARK:TRAC GAIN
Assigns marker 1 to the gain trace.
Characteristics
RST value: NOIS
SCPI: device–specific
Mode
NF
CALCulate:MARKer:X
This command positions the marker1 to the indicated frequency measurement point. If the
marker is off, it is switched on first.
This command is available from firmware version 1.60.
Parameter
0 to MAX (frequency | sweep time)
Example
CALC:MARK:X 550 MHZ
Positions marker 1 to frequency 550 MHz.
Characteristics
RST value: –
SCPI: device–specific
Mode
NF
CALCulate:MARKer:Y?
This command queries the measured value of the marker 1. If the marker is off, it is switched on
first. To obtain a valid query result, a complete sweep with synchronization to the sweep end
must be performed before the query of the y value. The query result is output in the unit
determined with the selected trace (see also DISPlay:DATA:TRACe<1> command).
This command is only a query and therefore has no *RST value.
This command is available from firmware version 1.60.
Example
CALC:MARK:Y?
Outputs the measured value of marker 1.