R&S FSL CALCulate Subsystem (Noise Figure, K30)
1300.2519.12 6.529 E-11
CALCulate:MARKer Subsystem (Noise Figure, K30)
The CALCulate:MARKer subsystem checks the marker functions of the instrument.
Commands of the CALCulate:MARKer Subsystem
– CALCulate:MARKer[:STATe]
– CALCulate:MARKer:AOFF
– CALCulate:MARKer:TRACe
– CALCulate:MARKer:X
– CALCulate:MARKer:Y?
– CALCulate:MARKer:MAXimum[:PEAK]
– CALCulate:MARKer:MiNimum[:PEAK]
CALCulate:MARKer[:STATe]
This command switches on or off the marker 1 for the selected trace. It is only available if
measurement results are displayed.
If the selected trace is switched off (DISPlay[:WINDow<1>]:TRACe<1|2>[:STATe]), the
marker is also switched off. If the marker is switched on again, the marker is set to the active
trace. If both noise and gain traces are switched off, the marker can not be switched on again.
This command is available from firmware version 1.60.
Parameter
ON | OFF
Example
CALC:MARK ON
Marker 1 is switched on.
Characteristics
RST value: OFF
SCPI: device–specific
Mode
NF
CALCulate:MARKer:AOFF
This command switches off the marker 1.
This command is an event and therefore has no *RST value and no query.
This command is available from firmware version 1.60.
Example
CALC:MARK:AOFF
Marker 1 is switched off.
Characteristics
RST value: –
SCPI: device–specific
Mode
NF