R&S FSL SENSe Subsystem (WLAN, K91 / K91n)
1300.2519.12 6.731 E-11
[SENSe:]DEMod:FILTer:MODulation
This command selects the TX and RX filters. The names of the filters correspond to the file
names; a query of all available filters is possible by means of the
[SENSe:]DEMod:FILTer:CATalog? command.
Parameter
<string>,<string>
DEF_TX: default transmit filter, DEF_RX: default receive filter
Example
DEM:FILT:MOD 'DEF_TX','DEF_RX'
DEF_TX is selected for the TX filter and DEF_RX for the RX filter
Characteristics
*RST value: AUTO,AUTO
SCPI: device–specific
Mode
WLAN
[SENSe:]DEMod:FORMat[:BCONtent]:AUTo
When this command is set to ON then the signal symbol field of the burst is analyzed to
determine the modulation scheme of the first burst. When this field is set to ON only bursts
which match the modulation scheme are considered in results analysis.
Parameter
ON | OFF
Example
DEM:FORM:AUT ON
Specifies that the signal symbol field should be decoded.
Characteristics
*RST value: –
SCPI: device–specific
Mode
WLAN
[SENSe:]DEMod:FORMat:BANalyze
The remote control command sets the analysis modulation format that will be assumed when
the measurement is performed. If the [SENSe:]DEMod:FORMat:SIGSymbol parameter has
been set to ON, this command can be used to measure only certain burst types within a
measurement sequence.
Parameter
For details refer to "Analysis modulation format" on page 6.727.
Example
DEM:FORM:BAN 'QAM16'
Only bursts that are of the QAM16 modulation format are analyzed.

Table of Contents