SENSe Subsystem (WLAN, K91 / K91n) R&S FSL
1300.2519.12 6.734 E-11
[
SENSe:]DEMod:FORMat:BANalyze:SYMBols:EQUal
When this command is activated then only bursts of equal length will take part in the
measurement analysis. When this command is set to true the value specified by the
[SENSe:]DEMod:FORMat:BANalyze:SYMBols:MIN command specifies the number of
symbols that a burst must have in order to take part in analysis.
Parameter
ON | OFF
Example
DEM:FORM:BAN:SYM:EQU ON
Only bursts of equal length will take part in analysis.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
WLAN
[SENSe:]DEMod:FORMat:BANalyze:SYMBols:MAX
This command specifies the maximum number of data symbols required for bursts to qualify for
measurement analysis. Only bursts with the specified number of symbols will be used in the
measurement analysis. The number of data symbols is defined as the uncoded bits including
service and tail bits.
This value will not have any immediate effect if the
[SENSe:]DEMod:FORMat:BANalyze:SYMBols:EQUal command has been set to true as in
this case no range of symbols is allowed and only bursts with exactly the number of symbols
specified by the [SENSe:]DEMod:FORMat:BANalyze:SYMBols:MIN command shall take
place in measurement analysis.
Parameter
<numeric_value>
Example
DEM:FORM:BAN:SYM:MAX 1300
Only bursts which contain a maximum symbol count of 1300 are analyzed.
Characteristics
*RST value: 64
SCPI: device–specific
Mode
WLAN