Configuring the Sweep Mode – SWEEP Key R&S FSL
1300.2519.12 4.26 E-11
Continue Single Sweep
Repeats the number of sweeps set by using the Sweep Count softkey, without deleting the
trace of the last measurement.
This is particularly of interest when using the trace configurations Average or Max Hold to take
previously recorded measurements into account for averaging / maximum search. For details on
trace configuration refer to "Setting Traces – TRACE Key" on page 4.39.
Remote: INIT:CONM
Sweeptime Manual
Opens an edit dialog box to enter the sweep time. For details see Sweeptime Manual softkey in
the bandwidth menu.
Remote: SWE:TIME 10s
Sweeptime Auto
Sets the automatic sweep time mode. For details see Sweeptime Auto softkey in the bandwidth
menu.
Remote: SWE:TIME:AUTO ON
Sweep Count
Opens an edit dialog box to enter the number of sweeps to be performed in the single sweep
mode. Values from 0 to 32767 are allowed. If the values 0 or 1 are set, one sweep is performed.
The sweep count is applied to all the traces in a diagram.
The sweep count set in the sweep menu is the same as that in the trace menu (for further details
see Sweep Count softkey). If the trace configurations Average, Max Hold or Min Hold are set,
the sweep count value also determines the number of averaging or maximum search
procedures (for details on trace configuration see "Setting Traces – TRACE Key" on page 4.39.
Example:
TRACE key – Trace Mode softkeyMode Max Hold softkey
SWEEP key – Sweep Count softkey – Average Sweep Count dialog box: enter 10
Single Sweep softkey: R&S FSL performs the Max Hold function over 10 sweeps.
Remote: SWE:COUN 64