Using Markers and Delta Markers – MKR Key R&S FSL
1300.2519.12 4.64 E-11
ASCII File Export
Opens the ASCII File Export Name dialog box and saves the content of the marker peak list in
ASCII format to the specified file and directory. The file consists of a data section containing the
peak list.
Example:
Peak;1
1089743590;Hz
–105.24;dBm
...
This format can be processed by spreadsheet calculation programs, e.g. MS Excel. It is
necessary to define ';' as a separator for the data import. Different language versions of
evaluation programs may require a different handling of the decimal point. It is therefore possible
to select between separators '.' (decimal point) and ',' (comma) using the Decim Sep softkey.
This softkey is available from firmware version 1.80.
Remote: FORM ASC
Remote: MMEM:STOR:PEAK 'test'
Decim Sep
For details refer to the Decim Sep softkey in the trace menu of the base unit.
This softkey is available from firmware version 1.80.
Marker Stepsize
Opens a submenu to set the step size of all markers and delta markers.
This softkey is available from firmware version 1.60.
Stepsize Standard
Moves the marker or delta marker from one measurement point to the next, if the marker or
delta marker value is changed via the rotary knob (Marker 1/Marker 2/Marker 3/Marker 4
softkeys). If more measured values than measurement points exist, it is not possible to read out
all measured values. In this case, use the Stepsize Sweep Points softkey.
This softkey is available from firmware version 1.60.
Remote: CALC:MARK:X:SSIZ STAN