Power Measurements – MEAS Key R&S FSL
1300.2519.12 4.88 E-11
Child Node Attribute Value Parameter Description Mandatory
node is used
RF Atennuation Mode Manual | Auto RF Attenuator Yes, if the ReferenceLevel
child node is used
Preamplifier ON | OFF Preamp Yes
ASCII file export format (Spectrum Emission Mask)
The first part of the file lists information about the spectrum analyzer and the general setup. For a
detailed description refer to "ASCII file export format" on page 4.51.
File contents Description
RefType; CPOWER;
TxBandwidth;9540000;Hz
Filter State; ON;
Alpha;0.22;
reference range setup, for details see
Edit Reference Range softkey
PeaksPerRange;1;
Values;4;
evaluation list information
0;–22500000;–9270000;1000000;2986455000;–74.762840270996094;
–10.576210021972656;–45.762840270996094;PASS;
1;–9270000;–4770000;100000;2991405000;–100.17695617675781;
–35.990325927734375;–1.490325927734375;PASS
3;4770000;9270000;100000;3005445000;–100.17695617675781;
–35.990325927734375;–1.490325927734375;PASS;
4;9270000;22500000;1000000;3018225000;–74.762840270996094;
–10.576210021972656;–45.762840270996094;PASS;
information about each peak:
<range number>;
<start frequency>;
<stop frequency>;
<resolution bandwidth of range>;
<frequency of peak>;
<absolute power in dBm of peak>;
<relative power in dBc of peak
(related to the channel power)>;
<distance to the limit line in dB
(positive value means above the limit)>;
<limit fail (pass = 0, fail =1)>;
Softkeys of the power measurement menu
The following table shows all softkeys available in the power measurement menu. It is possible that
your instrument configuration does not provide all softkeys. If a softkey is only available with a special
option, model or (measurement) mode, this information is delivered in the corresponding softkey
description.
Menu / Command Submenu / Command Submenu / Command Command
All Functions Off
TOI Marker 1
Marker 2
Marker 3
Marker 4
Search Signals
AM Mod Depth same contents as TOI
menu
Time Domain Power Peak
RMS
Mean