EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 272

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL Power Measurements – MEAS Key
1300.2519.12 4.87 E-11
Table 4-7: Attributes and child nodes of the Range element (normal ranges)
Child Node Attribute Value Parameter Description Mandatory
Index 0…19 Indexes are continous and
have to start with 0
Yes
Name <string> Name of the range Only if ReferenceChannel
contains a name and the
range is the reference
range
ShortName <string> Short name of the range No
ChannelType TX | Adjacent Yes
WeightingFilter Only if ReferencePower
method is “TX Channel
Power” and the range is
the reference range
Type RRC | CFILter Type of the weighting filter Yes
RollOffFactor 0…1 Excess bandwidth of the filter Only if the filter type is
RRC
Bandwidth <Bandwidth in Hz> Filter bandwidth Only if the filter type is
RRC
FrequencyRange Yes
Start <frequency in Hz> Start value of the range Yes
Stop <frequency in Hz> Stop value of the range Yes
Limit A Range must contain exactly
two limit nodes; one of the
limit nodes has to have a
relative unit (e.g. dBc), the
other one must have an
absolute unit (e.g. dBm)
Yes
Start Value <numeric_value> Power limit at start frequency Yes
Unit dBm/Hz | dBm | dBc |
dBr | dB
Sets the unit of the start value Yes
Stop Value <numeric_value> Power limit at stop frequency Yes
Unit dBm/Hz | dBm | dBc |
dBr | dB
Sets the unit of the stop value Yes
LimitFailMode Absolute | Relative |
Absolute and Relative |
Absolute or Relative
No (if quoted, it has to be
equal to
DefaultLimitFailMode)
RBW Bandwidth <bandwidth in Hz> RBW Yes
Type NORM | PULS | CFIL |
RRC
No
VBW Bandwidth <bandwidth in Hz> VBW Yes
Detector NEG | POS | SAMP |
RMS | AVER | QUAS
Detector No (if quoted, it has to be
equal in all ranges)
Sweep Mode Manual | Auto Sweep Time Mode Yes
Time <time in sec> Sweep Time No
Amplitude No
ReferenceLevel Value <power in dBm> Ref. Level Yes, if the ReferenceLevel
child node is used
Unit dBm Yes, if the ReferenceLevel

Table of Contents

Related product manuals