R&S FSL Bluetooth Measurements (Option K8)
1300.2519.12 4.159 E-11
Measurements overview
The Bluetooth Measurements option provides the following measurement types:
– "Output Power" on page 4.161
– "Adjacent Channel Power" on page 4.162
– "Modulation Characteristics" on page 4.162
– "Initial Carrier Frequency Tolerance" on page 4.163
– "Carrier Frequency Drift" on page 4.164
– "Relative Transmit Power (EDR)" on page 4.164
– "In–band Spurious Emissions (EDR)" on page 4.165
– "Carrier Frequency Stability and Modulation Accuracy (EDR)" on page 4.166
– "Differential Phase Encoding (EDR)" on page 4.166
The basic parameter settings are described in section "To adapt the settings to the characteristics of the
DUT" on page 4.159. The settings that can be configured individually for each measurement are the
following:
• RBW (the IF bandwidth set up for modulation measurements is valid for all measurements)
• VBW
• RBW auto mode
• VBW auto mode
• trace mode
• detector
• sweep count
• sweep time auto mode
• sweep time
They are available in the corresponding menus as soon as the corresponding measurement is selected.
Changes to these settings are always related to the selected measurement. The settings defined in the
RF Test Specification can thus be modified individually for development or production. By using the
start recall function, the individual configuration of the various tests can be preserved over a preset.
To adapt the settings to the characteristics of the DUT
1. Set the spectrum analyzer to its default state.
– Press PRESET.
The R&S FSL is set to its default state.
2. Select the Bluetooth operating mode.
– Press MODE key.
– In the Measurement Modes dialog box, select Bluetooth.
The Bluetooth mode is activated and the main menu of the option is displayed.
3. Select the transmit channel.
– Press Channel softkey and enter the desired channel number.