R&S FSL Bluetooth Measurements (Option K8)
1300.2519.12 4.179 E-11
Power Avg Stop (Output Power)
Opens an edit dialog box to enter the stop position of the evaluation area for the average burst
power. The values range from 0 to 100%, the default setting is 80%.
For further information refer to "Position of a Bluetooth burst" on page 4.160.
Remote: CONF:BTO:POW:AVER:STAR 90PCT
Zoom (Modulation Char, Init Carr Freq Tol, Carr Freq Drift)
Activates or deactivates the zoom function. Opens an edit dialog box to enter the zoom start
position. The valid value range for the zoom start position is 0 to (measurement time – 500 /
sampling rate). The default setting for the zoom function is 0s (off).
With active zoom function, an area of only 501 samples is displayed.
Remote: ADEM:ZOOM ON
Remote: ADEM:ZOOM:STAR 500us
No. of ACP Chan (TX Spec ACP, Spurious Emissions)
Opens an edit dialog box to enter the number of adjacent channels, for which the power is to be
measured. The values range from 0 to 78. A minimum value of 3 is recommended. The default
setting is 78 (all channels).
As with the ACP measurement of the basic instrument firmware this value refers to the number
of adjacent channels on one side of the TX channel. This means that with a selected value of 10
the analyzer will measure in total 21 channels (10 lower channels + TX channel + 10 upper
channels).
The frequency range required for the measurement is set up automatically. The center
frequency will also be adapted automatically dependent on the selected TX channel.
The measurement of the adjacent channels is limited to the available Bluetooth frequency band,
which means that at maximum 79 channels (23 channels in France) will be measured.
Remote: CONF:BTO:ACLR:ACP 10 (TX Spec ACP)
Remote: CONF:BTO:IBS:ACP 20 (Spurious Emissions)
Channel List Start (TX Spec ACP, Spurious Emissions)
Opens an edit dialog box to enter the channel with which the result display table starts. Because
of the multiplicity of the measurements results not all channels can be displayed simultaneously.
Gate Delay (Spurious Emissions)
Opens an edit dialog box to enter the time between trigger event and start of the DPSK packet.
That is the measurement start time.
Remote: SWE:EGAT:HOLD 100us