R&S FSL Noise Figure Measurements Option (K30)
1300.2519.12 4.265 E-11
Softkeys of the sweep menu (Noise mode)
The following table shows all softkeys available in the sweep menu in Noise mode (SWEEP key). It is
possible that your instrument configuration does not provide all softkeys. If a softkey is only available
with a special option, model or (measurement) mode, this information is delivered in the corresponding
softkey description.
Command
Sweep Single/Cont
Cal
Fix Freq
Sweep Single/Cont
Selects the sweep mode.
Single
single sweep mode
Cont
continuous sweep mode
If a measurement is started while another measurement is in progress, the first measurement
will be aborted and the new measurement started immediately.
For further details refer to "Measurement forms" on page 4.249.
Remote: CONF:LIST:CONT
Remote: CONF:LIST:SING
Cal
Performs a calibration. The calibration status of the noise figure measurement is displayed in the
title bar. For further details refer to "Calibration" on page 4.248.
This softkey is only available, if the 2nd Stage Correction option in the Measurement Settings
dialog box is activated.
Remote: CONF:CORR
Fix Freq
Starts a fixed frequency measurement for the frequency that is currently selected in the
Frequency List Results. For further details refer to "Measurement forms" on page 4.249 and
"Result displays" on page 4.250.
This softkey is only available after a frequency list measurement has been completed and the
measurement results are displayed in list form (Display List/Graph softkey).
Remote: CONF:SING
Remote: FREQ 10MHz