R&S FSL Noise Figure Measurements Option (K30)
1300.2519.12 4.267 E-11
Data –> Mem1/Data –> Mem2/Data –> Mem3
Saves the current trace results to trace memory <n>. If a trace memory <n> contains data, the
corresponding softkey has a green background. The content of the trace memory <n> is
displayed via the Mem1 On/Off/Mem2 On/Off/Mem3 On/Off softkeys.
If data is transferred to a trace memory that already contains trace data, the new trace data
overwrite the current trace data in the memory.
Remote: CONF:ARR:MEM2 ONCE
Remote: FETC:ARR:MEM2:NOIS:FIG?
Remote: FETC:ARR:MEM2:NOIS:GAIN?
Remote: FETC:ARR:MEM2:NOIS:TEMP?
Data On/Off
Switches the display of the current measurement results on or off. The display of trace memory
results is not affected if this softkey is pressed. If a new frequency list measurement is started,
the display of the current result trace is switched on automatically.
Remote: DISP:CURR:DATA OFF
Mem1 On/Off / Mem2 On/Off / Mem3 On/Off
Switches the display of trace memory <n> on or off. This softkey is not available if no data is
held in the selected trace memory.
Remote: DISP:ARR:MEM2 ON
Softkeys of the marker menu (Noise mode)
The following table shows all softkeys available in the marker menu in Noise mode (MKR key). It is
possible that your instrument configuration does not provide all softkeys. If a softkey is only available
with a special option, model or (measurement) mode, this information is delivered in the corresponding
softkey description.
Command
Marker 1
Marker to Trace
All Marker Off
Marker 1
Activates marker 1 and opens an edit dialog box to enter a value for marker 1 to be set to.
Pressing the softkey again deactivates the marker 1. This softkey is only available if
measurement results are displayed.
This softkey is available from firmware version 1.60.
Remote: CALC:MARK ON
Remote: CALC:MARK:X 550 MHZ
Remote: CALC:MARK:Y?