R&S FSL CALCulate Subsystem
1300.2519.12 6.43 E-11
CALCulate<1|2>:LIMit<1...8>:CONTrol:SHIFt
This command moves a limit line by the indicated value in x direction. In contrast to
CALCulate<1|2>:LIMit<1...8>:CONTrol:OFFSet, the line is shifted by modifying the
individual x values and not by means of an additive offset.
In analyzer mode, the unit of values depends on the span setting of the x–axis. For details refer
to CALCulate<1|2>:LIMit<1...8>:CONTrol:DOMain.
T
he numeric suffixes <1|2> are irrelevant for this command.
This command is an event and therefore has no *RST value and no query.
This command is available from firmware version 1.10.
Parameter
<numeric_value>
Example
CALC:LIM2:CONT:SHIF 50KHZ
Shifts all reference values of limit line 2 by 50 kHz.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CDMA, EVDO, NF
CALCulate<1|2>:LIMit<1...8>:CONTrol:SPACing
This command selects linear or logarithmic interpolation for the calculation of limit lines from
frequency points.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
LINear | LOGarithmic
Example
CALC:LIM:CONT:SPAC LIN
Characteristics
*RST value: LIN
SCPI: device–specific
Mode
A, CDMA, EVDO