R&S FSL CALCulate Subsystem
1300.2519.12 6.41 E-11
CALCulate:LIMit:CONTrol Subsystem
The CALCulate:LIMit:CONTrol subsystem defines the x–axis (CONTrol axis).
This subsystem is available from firmware version 1.10.
Commands of the CALCulate:LIMit:CONTrol Subsystem
CALCulate<1|2>:LIMit<1...8>:CONTrol[:DATA]
CALCulate<1|2>:LIMit<1...8>:CONTrol:DOMain
CALCulate<1|2>:LIMit<1...8>:CONTrol:MODE
CALCulate<1|2>:LIMit<1...8>:CONTrol:OFFSet
CALCulate<1|2>:LIMit<1...8>:CONTrol:SHIFt
CALCulate<1|2>:LIMit<1...8>:CONTrol:SPACing
CALCulate<1|2>:LIMit<1...8>:CONTrol[:DATA]
This command defines the x–axis values (frequencies or times) of the upper or lower limit lines.
The number of values for the CONTrol axis and for the corresponding UPPer and/or LOWer
limit lines has to be identical. Otherwise default values are entered for missing values or not
required values are deleted.
In analyzer mode, the unit of values depends on the span setting of the x–axis. For details refer
to CALCulate<1|2>:LIMit<1...8>:CONTrol:DOMain.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
<numeric_value>,<numeric_value>
Example
CALC:LIM2:CONT 1MHz,30MHz,100MHz,300MHz,1GHz
Defines 5 reference values for the x–axis of limit line 2.
CALC:LIM2:CONT?
Outputs the reference values for the x–axis of limit line 2 separated by a comma.
Characteristics
*RST value: – (CALCulate<1|2>:LIMit<1...8>:STATe is set to OFF)
SCPI: conform
Mode
A, ADEMOD, CDMA, EVDO, NF
CALCulate<1|2>:LIMit<1...8>:CONTrol:DOMain
This command defines the span setting for the x–axis values.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
FREQuency | TIME

Table of Contents