R&S FSL CALCulate Subsystem
1300.2519.12 6.89 E-11
P
arameter
ACPower
Adjacent–channel power measurement
Results are output in the following sequence, separated by commas:
Power of transmission channel
Power of lower adjacent channel
Power of upper adjacent channel
Power of lower alternate channel 1
Power of upper alternate channel 1
Power of lower alternate channel 2
Power of upper alternate channel 2
The number of measured values returned depends on the number of
adjacent/alternate channels selected with
[SENSe<1|2>:]POWer:ACHannel:ACPairs.
With logarithmic scaling (RANGE LOG), the power is output in the currently
selected level unit; with linear scaling (RANGE LIN dB or LIN %), the
power is output in W. If [SENSe<1|2>:]POWer:ACHannel:MODE is set
to REL, the adjacent/alternate–channel power is output in dB.
CN Measurement of carrier–to–noise ratio
The carrier–to–noise ratio in dB is returned.
CN0 Measurement of carrier–to–noise ratio referenced to 1 Hz bandwidth.
The carrier–to–noise ratio in dB/Hz is returned.
CPOWer Channel power measurement
In a Spectrum Emission Mask measurement, the query returns the power
result for the reference range, if this power reference type is selected.
With logarithmic scaling (RANGE LOG), the channel power is output in the
currently selected level unit; with linear scaling (RANGE LIN dB or LIN %),
the channel power is output in W.
MCACpower Channel/adjacent–channel power measurement with several carrier
signals
Results are output in the following sequence, separated by commas:
Power of carrier signal 1 to 12 in ascending order
Total power of all carrier signals
Power of lower adjacent channel
Power of upper adjacent channel
Power of lower alternate channel 1
Power of upper alternate channel 1
Power of lower alternate channel 2
Power of upper alternate channel 2
The number of measured values returned depends on the number of
carrier signals and adjacent/alternate channels selected with
[SENSe<1|2>:]POWer:ACHannel:TXCHannel:COUNt and
[SENSe<1|2>:]POWer:ACHannel:ACPairs.
If only one carrier signal is measured, the total value of all carrier signals
will not be output.
With logarithmic scaling (RANGE LOG), the power is output in dBm; with
linear scaling (RANGE LIN dB or LIN %), the power is output in W. If
[SENSe<1|2>:]POWer:ACHannel:MODE is set to REL, the
adjacent/alternate–channel power is output in dB.