CALCulate Subsystem R&S FSL
1300.2519.12 6.88 E-11
C
ALCulate<1|2>:MARKer:FUNCtion:POWer:PRESet
This command selects the power measurement setting for a standard and previously switches
on the corresponding measurement, if required.
The configuration for a standard comprises of the parameters weighting filter, channel
bandwidth and spacing, resolution and video bandwidth, as well as detector and sweep time.
The numeric suffixes <1|2> are not relevant.
Note: The settings for standards IS95A and C differ as far as the calculation method of
channel spacings is concerned. For IS95A and J–STD008 the spacing is calculated
from the center of the main channel to the center of the corresponding adjacent
channel, for IS95C from the center of the main channel to the nearest border of the
adjacent channel.
Parameter
NADC | TETRA | PDC | PHS | CDPD | FWCDma | RWCDma | F8CDma | R8CDma | F19Cdma |
R19Cdma | FW3Gppcdma | RW3Gppcdma | S2CDma | FIS95A | RIS95A | FIS95C0 | RIS95C0
| FIS95C1 | RIS95C1 | FJ008 | RJ008 | TCDMa | FTCDMa | RTCDMa | AWLan | BWLan |
WIMax | WIBRo | RFID14443 | NONE
For further details refer to "Predefined CP / ACP standards" on page 6.86.
The parameters WIMax and WIBRo are available from firmware version 1.30.
Example
CALC:MARK:FUNC:POW:PRES NADC
Selects the standard setting for NADC
Characteristics
*RST value: –
SCPI: device–specific
Mode
A–F
CALCulate<1|2>:MARKer:FUNCtion:POWer:RESult?
This command queries the result of the performed power measurement. If necessary, the
measurement is switched on prior to the query.
The channel spacings and channel bandwidths are configured in the "SENSe:POWer
Subsystem" on page 6.220.
To obtain a correct result, a complete sweep with synchronization to the end of the sweep must
be performed before a query is output. Synchronization is possible only in the single sweep
mode.
The numeric suffixes <1|2> are not relevant.
This command is a query and therefore has no *RST value.