SENSe Subsystem R&S FSL
1300.2519.12 6.202 E-11
C
haracteristics
*RST value: SPAN
SCPI: device–specific
M
ode
A, ADEMOD, CDMA, EVDO
[SENSe<1|2>:]FREQuency:CENTer:STEP:LINK:FACTor
This command couples the step size of the center frequency with a factor to the span (span >0)
or to the resolution bandwidth (span = 0).
The numeric suffixes <1|2> are not relevant.
Parameter
1 to 100 PCT
Example
FREQ:CENT:STEP:LINK:FACT 20PCT
Characteristics
*RST value: – (AUTO 0.1 × SPAN is switched on)
SCPI: device–specific
Mode
A, CDMA, EVDO
[SENSe<1|2>:]FREQuency:MODE
This command switches between span > 0 (SWEep) and zero span (CW | FIXed) in the
Spectrum Analyzer mode.
For CW and FIXed, the frequency setting is via the [SENSe<1|2>:]FREQuency:CENTer
command. In the sweep mode, the setting is via the [SENSe<1|2>:]FREQuency:STARt,
[SENSe<1|2>:]FREQuency:STOP, [SENSe<1|2>:]FREQuency:CENTer, and
[SENSe<1|2>:]FREQuency:SPAN commands.
The numeric suffixes <1|2> are not relevant.
Parameter
CW | FIXed | SWEep | SCAN
Example
FREQ:MODE SWE
Characteristics
*RST value: SWEep
SCPI: conform
Mode
A, CDMA