STL receiver handles position estimation. The default setting is recom-
mended for most applications.
Sensitivity Level: [default = 40] This value determines the sensitivity
of the STL receiver towards the STL signal bursts transmitted by the
satellites. The lower the number, the more responsive the receiver will
be to acknowledge the bursts. The default value is optimized for an
indoor antenna. A higher value can be used for outdoor antenna
installations (not typical). A value lower than (40) is not recom-
mended.
3.
Assign the STL 0 reference a reference priority by navigating to
MANAGEMENT > OTHER: Reference Priority.
4.
In the Configure Reference Priorities panel, click the + icon in the upper
right corner. The Add Reference window will open.
5.
Select a Priority Level:
a.
If STL is to be used as a backup to GPS: Select a Priority Level of 2.
b.
If STL is the only reference: Select a Priority Level of 1.
For Time and PPS, select STL 0. Click Submit.
6.
Verify your settings in the Reference Priority table, and ensure that the
new reference is Enabled.
Reviewing the STL Status
Validity Status
To check or monitor the validity of the STL reference:
1.
Navigate to INTERFACES > REFERENCES.
2.
In the References status panel, under STL 0, check the status indicator