R&S FSL SENSe Subsystem (CDMA, K82)
1300.2519.12 6.621 E-11
Example
INST:SEL BC2K
Activates the CDMA2000 BTS Analyzer option.
INIT:CONT OFF
Selects single sweep.
CDP:SFAC 128
Selects base spreading factor 128.
INIT;*WAI
Starts measurement with synchronization.
Characteristics
RST value: 64
SCPI: device-specific
Mode
CDMA
[SENSe<1|2>:]CDPower:SLOT
This command selects the power control group (PCG).
The numeric suffix at SENSe<1|2> is irrelevant for this command.
Parameter
0 to (capture length–1)
The capture length is defined via the [SENSe<1|2>:]CDPower:IQLength command.
Example
INST:SEL BC2K
Activates the CDMA2000 BTS Analyzer option.
INIT:CONT OFF
Selects single sweep.
CDP:SLOT 2
Selects power control group 2.
INIT;*WAI
Starts measurement with synchronization.
Characteristics
RST value: 0
SCPI: device-specific
Mode
CDMA
[SENSe<1|2>:]CDPower:TPMeas
This command actives or deactivates the timing and phase offset evaluation of the channels to
the pilot.
The numeric suffix at SENSe<1|2> is irrelevant for this command.
The results are queried using the TRACe<1|2>[:DATA] command, Channel Table return
values, and the CALCulate<1|2>:MARKer<1...4>:FUNCtion:CDPower[:BTS]:RESult?
command.
Parameter
ON | OFF