R&S FSL SENSe Subsystem (WLAN, K91 / K91n)
1300.2519.12 6.729 E-11
[
SENSe:]DEMod:BANalyze:DURation:EQUal
When this command is set to ON then only bursts of equal length will take part in the PVT
analysis. When this command is set to true the value specified by the
[SENSe:]DEMod:BANalyze:DURation:MIN command specifies the duration that a burst
must last in order to take part in measurement analysis.
Parameter
ON | OFF
Example
DEM:BAN:DUR:EQU ON
Only bursts of equal length will take part in the measurement analysis.
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
WLAN
[SENSe:]DEMod:BANalyze:DURation:MAX
This command specifies the maximum duration in microseconds required for bursts to qualify for
measurement analysis. Only bursts with the specified duration will be used in the measurement
analysis.
This value will not have any immediate effect if the
[SENSe:]DEMod:BANalyze:DURation:EQUal command has been set to true as in this
case no range of durations is allowed and only bursts with exactly the duration specified by the
[SENSe:]DEMod:BANalyze:DURation:MIN command shall take part in measurement
analysis
Parameter
<numeric_value>
Example
DEM:BAN:DUR:MAX 1300
Only bursts which have a maximum duration of 1300 microseconds are analyzed.
Characteristics
*RST value: 5464
SCPI: device–specific
Mode
WLAN
[SENSe:]DEMod:BANalyze:DURation:MIN
This command specifies the duration in microseconds required for bursts to qualify for
measurement analysis. Only bursts with the specified duration will be used in the measurement
analysis. When the [SENSe:]DEMod:BANalyze:DURation:EQUal command has been set
to true then this command specifies the exact duration required for a burst to take part in
measurement analysis. When the [SENSe:]DEMod:FORMat:BANalyze:DBYTes:EQUal
command is set to false this command specifies the minimum duration required for a burst to
take part in measurement analysis.