R&S FSL Changing Settings via Markers – MKR–> Key
1300.2519.12 4.71 E-11
Peak
Sets the active marker/delta marker to the highest maximum of the trace.
If the Spectrogram Measurement option (K14) is activated, this softkey provides altered
functionality. For details refer to Peak softkey in the marker–> menu of this option.
Remote: CALC:MARK:MAX
Remote: CALC:DELT:MAX
Next Peak
Sets the active marker/delta marker to the next maximum of the selected trace according to the
mode selected using the Next Peak Mode < abs > softkey.
If the Spectrogram Measurement option (K14) is activated, this softkey provides altered
functionality. For details refer to Next Peak softkey in the marker–> menu of this option.
Next Peak Mode < abs >
Selects the mode of the Next Peak softkey. Three settings are available:
< Sets the active marker/delta marker to the next maximum left to the marker of the
selected trace.
abs Sets the active marker/delta marker to the next lower maximum of the selected
trace.
> Sets the active marker/delta marker to the next maximum right to the marker of the
selected trace.
If the Spectrogram Measurement option (K14) is activated, this softkey provides altered
functionality. For details refer to Next Peak X Search < abs > and Next Peak Y Search
up/abs/dn softkeys in the marker–> menu of this option.
Remote: CALC:MARK:MAX:LEFT (>)
Remote: CALC:DELT:MAX:LEFT (<)
Remote: CALC:MARK:MAX:RIGH (>)
Remote: CALC:DELT:MAX:RIGH (>)
Remote: CALC:MARK:MAX:NEXT (abs)
Remote: CALC:DELT:MAX:NEXT (abs)
Center =Mkr Freq (span > 0)
Sets the center frequency to the current marker or delta marker frequency. A signal can thus be
set to as center frequency, for example to examine it in detail with a smaller span.
Remote: CALC:MARK:FUNC:CENT
Ref Lvl =Mkr Lvl
Sets the reference level to the current marker level.
Remote: CALC:MARK:FUNC:REF