Spectrogram Measurement (Option K14) R&S FSL
1300.2519.12 4.196 E-11
Select Frame
Opens an edit dialog box to select the frame if no measurement is running. If the active marker
value is altered, the value of the selected frame is updated accordingly.
During a measurement, the value is fixed to 0 and this softkey is not available.
Remote: CALC:SPEC:FRAM:SEL –10
Sweeptime Manual
For details refer to the Sweeptime Manual softkey in the sweep menu of the base unit.
Sweeptime Auto
For details refer to the Sweeptime Auto softkey in the sweep menu of the base unit.
Sweep Count
For details refer to the Sweep Count softkey in the sweep menu of the base unit.
Sweep Points
For details refer to the Sweep Points softkey in the sweep menu of the base unit.
Continue Frame On/Off
Determines whether the results of the last measurement are deleted before starting a new
measurement.
On
Repeats the single sweep measurement without deleting the spectrogram results of
the last measurement. One of the following trace modes is to be used: Max Hold, Min
Hold, Average.
Off
Deletes the last measurement results before performing a single sweep
measurement.
This softkey takes effect in single sweep mode only.
Remote: CALC:SPEC:CONT ON
Frame Count
Opens an edit dialog box to enter the number of frames to be recorded in a single sweep
measurement, whereas the Sweep Count entry determines how many sweeps are combined
according to the selected trace mode (Max Hold or Min Hold) in one trace (frame). The
maximum number of frames is determined according to "Maximum number of frames" on page
4.193.
This softkey takes effect in single sweep mode only.
Remote: CALC:SPEC:FRAM:COUN 200